Environmental Test Chamber
Created with Pixso.

Fluorescent Ultraviolet Aging Chamber Lamp Exposure Weathering Accelerated ISO 4892

Fluorescent Ultraviolet Aging Chamber Lamp Exposure Weathering Accelerated ISO 4892

Brand Name: Sinuo
Model Number: SN882B
MOQ: 1
Price: Customized
Payment Terms: T/T
Supply Ability: 20 sets/month
Detail Information
Place of Origin:
China
Certification:
Calibration Certificate (Cost Additional)
Applied Standard:
ASTM_D 4329, ISO 4892-2 And Etc.
Inner Size:
1100*620*500 Mm(W×D×H)
Irradiation Temperature:
50~ 70℃
Condensing Temperature:
50℃
Light Source:
UVA Or UVB Lamps
Lamp Lifespan:
Around 1600 Hours
Lamp Power:
40W
Lamps Quantity:
8 PCS
Packaging Details:
Plywood case
Supply Ability:
20 sets/month
Highlight:

Fluorescent ultraviolet aging chamber

,

ISO 4892 Lamp ultraviolet aging chamber

,

Exposure UV Aging Test Chamber

Product Description

 

 

Chamber Type Fluorescent Ultraviolet Lamp UV Aging Test Chamber 

 

Complied Standards & Clauses


The UV Weathering Accelerated Ageing Test Chamber is designed and manufactured according ton the standard requirements of  ASTM G 154, ISO 4892-2, ISO 4892-3; ISO 11507; ASTMD 4329, and etc.

 

Equipment Overview


The UV Weatherproof Test Chamber can simulate the hazards caused by sunlight, rain, and dew. Adopt fluorescent UV lamps to simulate the effect of sunlight, use condensed moisture to simulate rain and dew, and place the tested material at a certain temperature to alternate between light and moisture test in the loop program. The test chamber takes few days or few months, which can reproduce hazards that occur outdoors. Hazard types include: fading, discoloration, loss of gloss, chalking, cracking, turbidity, bubbles, embrittlement, strength loss, and oxidation. The UV test data can help to select new materials and evaluate how changes in formulation affect the durability of new products. It is applicable widely in coatings, inks, paints, resins, plastics, rubber, printing and packaging, aluminum profiles, adhesives, automobiles, motorcycle industries, cosmetics, metals, electronics, electroplating, medicine, etc.

 

Technical Parameters


Inner chamber size 1100*620*500 mm(W×D×H)
Outer chamber size Approx. 1500*780*1800mm(W×D×H)
Irradiation temperature 50~ 70℃
Condensing temperature RT~60℃
Spray temperature Room temperature
Temperature uniformity ±2℃
Temperature fluctuation ±1℃
Temperature resolution 0.01℃
Relative humidity Condensation period≥85%RH; Irradiation period≤75%RH
Lamp wavelength

UVA-340, band 315-400nm, irradiance 0~0.9W/m2

Or UV-B 313, band 280--315nm, irradiance 0~0.9W/cm2

Quantity of lamps 8 PCS
Lamp power 40W/PC
Lamp brand Q-lab
Lamp lifespan Around 1600 hours
Distance between sample and lamp 100~300mm can be adjustable, according to standard requirements, the distance should be 50mm.
Lamp Central distance  70mm
Lamp spacing distance 35mm
Number of spray nozzle 4 pcs stainless steel nozzle ararnged in the top low
Water spray pressure 70~200Kpa adjustable
Test time 1~9999 H, M, S, the time can be preset 
Control system Independent research and development of touch screen and PLC controller
Inner chamber material SUS304 stainless steel plate
Outer chamber material Painted A3 plate
Radiometer/Sun Eye 1. Model: TEMI-UVI
2. Detection wavelength range: 200nm ~ 480nm.
3. Irradiance measurement range: 0.4W/m2~1.6W/m2
4. Operating temperature range: normal temperature---70 degrees
5. Photosensitive area: Φ12mm
6. Appearance size: 30*30*35mm
7. The illuminance sensor can measure and control the radiation energy in real time and feed it back to the controller. The controller adjusts the PID control in real time according to the set value to ensure that the set value and the actual value are consistent.
Equipment use conditions Please install the water pipe in advance, and use the 1/2-inch pipe interface; the floor drain or drain pipe interface should be reserved for the placement of the equipment. The drain pipe interface is level with the ground.
 
UV-A lamp irradiance distribution map

Fluorescent Ultraviolet Aging Chamber Lamp Exposure Weathering Accelerated ISO 4892 0
 
UV-B lamp irradiance distribution map

 
Fluorescent Ultraviolet Aging Chamber Lamp Exposure Weathering Accelerated ISO 4892 1