| Brand Name: | Sinuo |
| Model Number: | SNQC1015 |
| MOQ: | 1 |
| Price: | Customized |
| Payment Terms: | T/T |
| Supply Ability: | 20 sets/month |
Temperature Control Water Bath Constant Temperature Water Bath 750L
Complied Standards & Clauses
This constant water bath is required by IEC 60811-402:2012, IEC502-83, etc.
Equipment Overview
This equipment is a water tank type. The inner tank and the cover are made of stainless steel. The chassis is carefully made of high-quality cold-rolled steel plates. It adopts electric heating, intelligent temperature control, and digital display. A circulating pump can be selected to force circulation to ensure temperature uniformity and achieve constant temperature of liquids such as oil and water.
This equipment has the characteristics of beautiful appearance, corrosion resistance, stable temperature control, energy saving and practicality, convenient maintenance, and long service life. It is a supporting product for quality inspection agencies and wire and cable companies to measure the DC resistance and insulation resistance of wires and cables. It is a device specially designed for electrical performance testing of PVC plastic cables. It can be used in constant temperature water bath occasions specified in IEC 60811-402:2012, IEC502-83, etc. It is a plastic cable insulation test equipment.
Equipped with an insulation port, which allows wires and cables to be threaded.
Technical Parameters
1.Inner dimensions (L*W*H): 1500×500×1000mm
2.Temperature control range: room temperature ~ 100℃, user can preset the required test temperature from room temperature ~ 100℃, such as set at 70 ℃ or 90 ℃.
3.Temperature stability: <1.5℃
4.Electric heating power: 18KW
5.Insulation power: 5KW
6.Input voltage: AC 380V/50HZ
7.Temperature uniformity guarantee: high-precision instrument control, heat circulation pump circulation.
8.With Low water level alarm function, when there is lacking water, the heating tube will be powered off.
9.There are 12 holes on the cover, two of each 1,2,3,4,5,10 CM. Layout as follows:
![]()
10.Automatic water replenishment function.
11.LCD panel display with computer interface (485).
Usage: Please read the manual carefully before use. The user needs to be familiar with the instrument and the control panel, and use a megohmmeter to measure whether the power supply, plug, and grounding are poorly insulated or short-circuited. This device uses distilled water to heat the electrolyte at a constant temperature. River water and hard water are prohibited. If tap water is used, the thermostatic pot should be cleaned after each use to prevent the accumulation of scale on the heater from affecting the thermostatic sensitivity.
Details
![]()
![]()